<?xml version="1.0"?>
<Articles JournalTitle="Auditory and Vestibular Research">
  <Article>
    <Journal>
      <PublisherName>Tehran University of Medical Sciences</PublisherName>
      <JournalTitle>Auditory and Vestibular Research</JournalTitle>
      <Issn>2423-480X</Issn>
      <Volume>33</Volume>
      <Issue>4</Issue>
      <PubDate PubStatus="epublish">
        <Year>2024</Year>
        <Month>04</Month>
        <Day>14</Day>
      </PubDate>
    </Journal>
    <title locale="en_US">Does the Receiver Type Used in Receiver-in-Canal Hearing Aids Have an Effect on Real-Ear-to-Coupler Difference and Coupler Response for Flat Insertion Gain Values?</title>
    <FirstPage>313</FirstPage>
    <LastPage>321</LastPage>
    <AuthorList>
      <Author>
        <FirstName>Masoud</FirstName>
        <LastName>Bolandi</LastName>
        <affiliation locale="en_US">Department of Audiology, University of Social Welfare and Rehabilitation Science, Tehran, Iran</affiliation>
      </Author>
      <Author>
        <FirstName>Ahmad Reza</FirstName>
        <LastName>Nazeri</LastName>
        <affiliation locale="en_US">Department of Audiology, School of Rehabilitation, Shahid Beheshti University of Medical Sciences, Tehran, Iran</affiliation>
      </Author>
      <Author>
        <FirstName>Hamid</FirstName>
        <LastName>Jalilvand</LastName>
        <affiliation locale="en_US">Department of Audiology, School of Rehabilitation, Shahid Beheshti University of Medical Sciences, Tehran, Iran</affiliation>
      </Author>
      <Author>
        <FirstName>Alireza</FirstName>
        <LastName>Akbarzadeh Baghban</LastName>
        <affiliation locale="en_US">Proteomics Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran AND Department of Biostatistics, School of Allied Medical Sciences, Shahid Beheshti University of Medical Sciences, Tehran, Iran</affiliation>
      </Author>
    </AuthorList>
    <History>
      <PubDate PubStatus="received">
        <Year>2023</Year>
        <Month>12</Month>
        <Day>15</Day>
      </PubDate>
      <PubDate PubStatus="accepted">
        <Year>2024</Year>
        <Month>02</Month>
        <Day>10</Day>
      </PubDate>
    </History>
    <abstract locale="en_US">Background and Aim: Real-Ear-to-Coupler Difference (RECD) is affected by the type of transducer and its coupling method. This study aimed to investigate the effect of receiver type used in Receiver-in-Canal (RIC) hearing aids on RECD and Coupler Response for Flat Insertion Gain (CORFIG) obtained by the same coupling method.
 Methods: In this study, the right ears of 30 normal-hearing adults (25 men and 5 women, with a mean age of 28.13&#xB1;5.62 years) were studied. We used the RIC hearing aids with two standard and power receivers. The foam of the ER-3A insert earphone was used as a coupling system in both receivers. By using the Audiogram Direct test within Phonak&#x2019;s fitting software, 70 dB HL intensity was produced at different frequencies, once in the realear and again in the HA-1 coupler. By assessing the difference between real-ear and HA-1 coupler measurements, the RECD was first obtained. Then, the CORFIG was obtained at different frequencies by subtracting the Microphone Location Effect (MLE) and RECD from Real-Ear Unaided Gain (REUG).
 Results: The RECD and CORFIG values obtained from the standard receiver at all frequencies were significantly different from those from the power receiver (p&lt;0.05), where the use of the power receiver resulted in lower RECD and higher CORFIG. With the increase in frequency, the RECD increased. A small frequency changes of 50 Hz had a significant effect on RECD and CORFIG.
 Conclusion: The type of receiver used in RIC hearing aids has a significant effect on the RECD and CORFIG values.</abstract>
    <web_url>https://avr.tums.ac.ir/index.php/avr/article/view/1232</web_url>